Monitör - NST - SPO2
BIONICS
   

PALMCARE HAND PULSEOXIMETER

Display
Screen ; 70.7 X 38.8 LCD display
11 segment pulse amplitude bar
Maximum 72 hours trend function
Numeric data display ; High SAT, High Rate, Low SAT, Low Rate,
Current Rate, SAT
- Alarms On/Off

Range
Saturation ; 0~100%
Pulse rate ; 30~250bpm (beats per minute)

Limits
- Preset alarm limits

Adults Neonates
High SAT 100 SpO2% 95 SpO2%
Low SAT 85 SpO2% 80 SpO2%
High Rate 140 bpm 200 bpm
Low Rate 55 bpm 100 bpm

- Alarm limit range

High Low
SpO2 alarm range 50~100% & Off 50~99% & Off
(1% steps) (1% steps)
Pulse rate alarm 30~300 bpm & Off 30~300 bpm & Off
range

Saturation accuracy (%SpO2 plus/minus 1SD)

Plus/minus 2% @ 70~100%
Plus/minum 2 bpm or 2% (whichever is greater) @ 30~300 bpm

Power requirements
- 9V DC 4.5 Amps 100~240 VAC, 50/60Hz
- Less than 24W power consumption

Battery pack
- Lithium ion battery
- 4 hours operation when fully charged

Output
- RS-232C port ; Trend data (SpO2 and pulse rate, inverval)

Trend memory
- Saturation and pulse rate memories (72 hours maximum)

Physical dimension
- 45mm (H) X 190mm (W) X 88mm (D)
- Weight ; 286g, 346g (include battery)

Temperature
- Operating ; 50 F to 113 F (10 to +45 Centigrade)
- Storage/Transportation ; 14F to 122F (-10 to +50 Centigrade)

Humidity
- Operating ; 30~85% non condensing
- Storage/Transporation ; 20~95% non condensing

Features
- Monitors adult, pediatrics and neonates
- Plethysmographic waveforms display and up to 72 hours of
patient trends
- High q'lty LCD display
- Long battery life ; 4 hours
- Digital signal processing
- Auto gain control signal processing

İSTANBUL
ANKARA
İZMİR

İnönü Cad, Sümko Sitesi, A2 D4 Kozyatağı 34736 İSTANBUL Telefon:216.4648144    Faks: 216.4648140

2.Cadde, No: 38/10, Öveçler 06460 ANKARA
Telefon: 312.4729210     Faks: 312.4729211
275/11 Sokak No:11/2-A Görkem Sitesi Bornova İZMİR
Telefon: 232.3477661   Faks: 232.3477661
E-mail: info@vegagroup.com.tr  
© Right by Vega Medikal Ltd. Şti. 2006Erenna